S. Pridgeon
Clinical evaluation of a simple uroflowmeter for categorization of maximum urinary flow rate
Pridgeon, S.; Harding, C.; Newton, D.; Pickard, R.
Authors
Abstract
Objective: To evaluate the accuracy and diagnostic usefulness of a disposable flowmeter consisting of a plastic funnel with a spout divided into three chambers. Materials and Methods: Men with lower urinary tract symptoms (LUTS) voided sequentially into a standard flowmeter and the funnel device recording maximum flow rate (Q max ) and voided volume (V void ). The device was precalibrated such that filling of the bottom, middle and top chambers categorized maximum input flows as <10, 10-15 and > 15 ml s -1 respectively. Subjects who agreed to use the funnel device at home obtained readings of flow category and V void twice daily for seven days. Results: A single office reading in 46 men using the device showed good agreement with standard measurement of Q max for V void > 150 ml (Kappa = 0.68). All 14 men whose void reached the top chamber had standard Q max > 15 ml s -1 (PPV = 100%, NPV = 72%) whilst eight of 12 men whose void remained in the bottom chamber had standard Q max < 10 ml s -1 (PPV = 70%, NPV = 94%). During multiple home use by 14 men the device showed moderate repeatability (Kappa = 0.58) and correctly categorized Q max in comparison to standard measurement for 12 (87%) men. Conclusions: This study suggests that the device has sufficient accuracy and reliability for initial flow rate assessment in men with LUTS. The device can provide a single measurement or alternatively multiple home measurements to categorize men with Q max < 15 ml s -1 .
Citation
Pridgeon, S., Harding, C., Newton, D., & Pickard, R. (2007). Clinical evaluation of a simple uroflowmeter for categorization of maximum urinary flow rate. Indian Journal of Urology, 23(2), 114-118. https://doi.org/10.4103/0970-1591.32058
Journal Article Type | Article |
---|---|
Publication Date | Apr 1, 2007 |
Deposit Date | Jun 1, 2012 |
Journal | Indian Journal of Urology |
Print ISSN | 0970-1591 |
Electronic ISSN | 1998-3824 |
Publisher | Medknow Publications |
Peer Reviewed | Peer Reviewed |
Volume | 23 |
Issue | 2 |
Pages | 114-118 |
DOI | https://doi.org/10.4103/0970-1591.32058 |
Keywords | Bladder outlet obstruction, Urinary symptoms, Uroflowmetry. |
Public URL | https://durham-repository.worktribe.com/output/1564845 |
You might also like
What do we know so far about the research-teaching nexus in Initial Teacher Training?
(2023)
Journal Article
Developing Creative Teaching Skills in Pre-Service Teachers
(2022)
Journal Article
‘Allowing them to dream’: fostering creativity in mathematics undergraduates
(2022)
Journal Article
The Creative Sciences
(2021)
Journal Article
Fostering creative thinking in a digital world
(2020)
Journal Article
Downloadable Citations
About Durham Research Online (DRO)
Administrator e-mail: dro.admin@durham.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2025
Advanced Search