Compact, modular and in-plane AOSLO for high-resolution retinal imaging

Young, Laura K.; Morris, Timothy J.; Saunter, Christopher D.; Smithson, Hannah E.

Compact, modular and in-plane AOSLO for high-resolution retinal imaging Thumbnail


Authors

Laura K. Young

Christopher D. Saunter

Hannah E. Smithson



Abstract

The adaptive optics scanning laser ophthalmoscope (AOSLO) was first developed in 2002 and since then the technology has been adopted in several laboratories around the world, for both clinical and psychophysical research. There have been a few major design implementations of the AOSLO. The first used on-axis tilted spherical mirrors in a planar arrangement, and the second minimized the build up of astigmatism present in the first design by using a non-planar arrangement. Other designs have avoided astigmatism by using custom-made toroidal mirrors or by using lenses on-axis, rather than mirrors. We present a new design implementation for an AOSLO that maintains a planar optical alignment without the build up astigmatism using compact, reconfigurable modules based on an Offner relay system. We additionally use an off-the-shelf digital oscilloscope for data capture and custom-written Python code for generating and analyzing the retinal images. This design results in a compact system that is simple to align and, being composed of modular relays, has the potential for additional components to be added. We show that this system maintains diffraction-limited image quality across the field of view and that cones are clearly resolved in the central retina. The modular relay design is generally applicable to any system requiring one or more components in the pupil conjugate plane. This is likely to be useful for any point-scanned system, such as a standard scanning laser ophthalmoscope or non-ophthalmic confocal imaging system.

Citation

Young, L. K., Morris, T. J., Saunter, C. D., & Smithson, H. E. (2018). Compact, modular and in-plane AOSLO for high-resolution retinal imaging. Biomedical Optics Express, 9(9), 4275-4293. https://doi.org/10.1364/boe.9.004275

Journal Article Type Article
Acceptance Date May 15, 2018
Online Publication Date Aug 15, 2018
Publication Date Aug 15, 2018
Deposit Date Aug 29, 2018
Publicly Available Date Sep 20, 2018
Journal Biomedical Optics Express
Electronic ISSN 2156-7085
Publisher Optica
Peer Reviewed Peer Reviewed
Volume 9
Issue 9
Pages 4275-4293
DOI https://doi.org/10.1364/boe.9.004275
Public URL https://durham-repository.worktribe.com/output/1350451

Files

Published Journal Article (6.2 Mb)
PDF

Publisher Licence URL
http://creativecommons.org/licenses/by/4.0/

Copyright Statement
Published by The Optical Society under the terms of the Creative Commons Attribution 4.0 License. Further distribution of this work must maintain attribution to the author(s) and the published article's title, journal citation, and DOI.