Abstract
The restrained dynamic creep behaviors of two clinical bone cements, Palacos R-40 and CMW1 have been investigated at room temperature and body temperature. It was found that the two cements demonstrated significantly different creep deformations, with Palacos R-40 bone cement demonstrating higher creep strain than CMW1 bone cement at each loading cycle. For both cements, two stages of creep were identified with a higher creep rate during early cycling followed by a steady-state creep rate. The test temperature had a strong effect on the creep performance of the bone cements with higher creep rate observed at body temperature. The relationship between creep deformation and loading cycles can be expressed by single logarithmic model. The SEM examinations revealed that CMW1 bone cement is more sensitive to defects within the specimen especially to the defects at the edges of the specimen than Palacos R-40 bone cement. However, in the absence of micro-cracks or defects within the inner surface layer, the dynamic loading (at less than 10.6 MPa) is unlikely to produce micro-cracks in the CMW1 bone cement. The different behaviors between the two bone cements may be attributed to differences in chemical compositions and molecular weight distributions.
Citation
Liu, C., Green, S., Watkins, N., Gregg, P., & McCaskie, A. (2002). Creep behavior comparison of CMW1 and palacos R-40 clinical bone cements. Journal of Materials Science: Materials in Medicine, 13(11), 1021-1028. https://doi.org/10.1023/a%3A1020328218520
Journal Article Type | Article |
---|---|
Publication Date | 2002-11 |
Deposit Date | Apr 23, 2008 |
Journal | Journal of Materials Science: Materials in Medicine |
Print ISSN | 0957-4530 |
Electronic ISSN | 1573-4838 |
Publisher | Springer |
Peer Reviewed | Peer Reviewed |
Volume | 13 |
Issue | 11 |
Pages | 1021-1028 |
DOI | https://doi.org/10.1023/a%3A1020328218520 |
Keywords | Total hip replacement. |
Public URL | https://durham-repository.worktribe.com/output/1587554 |
You might also like
A re-appraisal of wear features of acetabular sockets using atomic force microscopy.
(2002)
Journal Article
A novel technique for the detailed size characterization of wear debris
(2000)
Journal Article
The wear of metal-on-metal total hip prostheses measured in a hipsimulator
(2001)
Journal Article
A preliminary hip joint simulator study of the migration of a cemented femoral stem
(2003)
Journal Article
Downloadable Citations
About Durham Research Online (DRO)
Administrator e-mail: dro.admin@durham.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2025
Advanced Search